
General
As
part of the New Approach to technical harmonization in the European
Union, directives have also been established for medical devices. These
directives specify conditions to be met before permission is granted to
apply the CE Mark and cover all medical devices and accessories.
A
medical device is any instrument, apparatus, appliance, material or
other article, whether used alone or in combination, including the
software necessary for its proper application intended by the
manufacturer to be used for human beings for the purpose of:
-
diagnosis,
prevention, monitoring, treatment or alleviation of disease,
-
diagnosis,
monitoring, treatment, alleviation of or compensation for an injury
or handicap,
-
investigation,
replacement or modification of the anatomy or of a physiological
process,
-
control
of conception
A
medical device does not achieve its principal intended action in or on
the human body by pharmacological, immunological or metabolic means, but
may be assisted in its function by such means.
Three Directives
have been drawn up.
Active
Implantable Medical Devices
(AIMD: 90/385/EEC)
This
directive covers all medical devices, which rely for its functioning on
a power supply and are left in the human body. Examples of AIMD's are
implantable cardiac pacemakers, implantable nerve stimulators and
implantable drug administration devices with or without catheter.
As of January 1,
1995, all AIMD's require a CE marking.
Medical
Devices
(93/42/EEC)
The
Medical Device Directive covers most other medical devices (active and
non-active) and their accessories that are not covered by the first or
the third directive. This concerns a large number of products ranging
from walking aids to prosthetic heart valves.
The
directive took effect on January 1, 1995. After the transition period,
which ends on June 14, 1998, all medical devices placed on the market or
put into service and covered by this directive must have a CE marking.
In Vitro
Diagnostic Medical Devices (98/79/EC)
This directive
covers any medical device, instrument, apparatus or system which is
intended to be used in-vitro for the examination of substances derived
from the human body.
This
directive is entered into force on 7 June 2000. There is a transition
period of 5 years (till 7 June 2005) for the placing of devices on the
EEA market and 2 years more (till 7 June 2007) for the putting into
service of devices complying with pre-existing regulations.
The
subsequent sections of this brochure will deal with the Medical Device
Directive (93/42/EEC) only
CLASSIFICATION
OF MEDICAL DEVICES
A
classification system has been developed for the Medical Device
Directive, dividing the related devices into four classes based on the
risk factor involved and the intended use to be made of the product.
This classification is based on the following principles:
1.
the duration of contact between the device and the patient
2.
the degree and manner of the device penetrating the human body
3.
the degree of the device's impact on the human anatomy.
Classification
Rules
Based
on the above-mentioned principles, 18 classification rules have been
developed; Annex IX to the Directive contains a list of these rules.
Rules
1 through 4 apply to devices that do not penetrate the human body
(non-invasive devices), whereas rules 5 through 8 apply to devices that do
penetrate the human body (invasive devices). A distinction is made
between the manner of penetration the body (through a natural or an
artificially created orifice) and the duration of contact between the
device and the patient.
For
powered devices (active devices) supplementary rules 9 through 12 are
also applicable. A distinction is made between therapeutic and
diagnostic devices, devices administering or removing medicines, devices
administering or exchanging energy to the human body from it, and other
devices.
Rules
13 through 18 are special rules for specifically defined devices used
for special applications.
Classification
According
to the classification rules given in Annex IX to the directive, devices
are classified as follows:
This
includes hospital beds, incontinence diapers, ordinary band-aids,
external splints, spectacle glasses, examination gloves, reusable
surgical tools, non-invasive electrodes etc.
This
includes catheters, ultrasound equipment, blood filtration equipment,
standard contact lenses, surgical gloves, dental fillings, hearing
aids etc.
This
includes haemodialysis equipment, standard intra-ocular lenses, drug
administration devices, anaesthetic apparatus, contact lens fluids,
blood bags, X-ray equipment etc.
- Class
III for high-risk devices
This
includes cardiac valves, neurological catheters, implants with a
biologically active coating etc.
ESSENTIAL
REQUIREMENTS
Each
medical device must, regardless of its classification, comply with the
essential requirements specified in Annex I of the directive. The
essential requirements consist of general requirements and a number of
more specific requirements.
The general
requirements are:
-
The
devices must be designed and manufactured in such a way that, when
used under the conditions and for the purposes intended, they will
not compromise the clinical condition or the safety of patients, or
the safety and health of users or, where applicable, other persons,
provided that any risks which may be associated with their use
constitute acceptable risks when weighed against the benefits to the
patients and are compatible with a high level of protection of
health and safety.
-
The
solutions adopted by the manufacturer for the design and
construction of the devices must conform to safety principles,
taking account of the generally acknowledged state of the art.
In
selecting the most appropriate solutions, the manufacturer must apply
the following principles in the following order:
-
eliminate
or reduce risks as far as possible (inherently safe design and
construction)
-
where
appropriate, take adequate protection measures including alarms, if
necessary, in relation to risks that cannot be eliminated
-
inform
users of the residual risks due to any shortcomings of the
protection measures adopted.
-
The
devices must achieve the performances intended by the manufacturer
and be designed, manufactured and packaged in such a way that they
are suitable for one or more of the functions referred to in Article
1 (2) (a), as specified by the manufacturer.
-
The
characteristics and performances referred to in Sections, 1, 2 and 3
must not be adversely affected to such a degree that the clinical
conditions and safety of the patients and, where applicable, of
other persons are compromised during the lifetime of the device as
indicated by the manufacturer, when the device is subjected to the
stresses which can occur during normal conditions of use.
-
The
devices must be designed, manufactured and packed in such a way that
their characteristics and performances during their intended use
will not be adversely affected during transport and storage, taking
account of the instructions and information provided by the
manufacturer.
-
Any
undesirable side-effect must constitute an acceptable risk when
weighed against the performances intended.
Specific
requirements relate to:
-
chemical,
physical and biological properties
-
infection
and microbial contamination
-
construction
and environmental properties
-
devices
with a measuring function
-
protection
against radiation
-
medical
devices connected to or equipped with an energy source
-
information
supplied by the manufacturer
-
clinical
data
PRODUCT
QUALITY ASSURANCE
Generally
speaking, the manufacturer's quality system should ensure that the
design and/or manufacture of a medical device is in compliance with the
requirements of the directive. If the relevant harmonized standards are
implemented in the quality system, it is presumed to conform to the
directive.
A
selection can be made of several quality assurance methods, depending on
the classification of the products and the corresponding conformity
procedure in accordance with the directive. The manufacturer can apply a
`full quality assurance system' or e.g. a `product quality assurance
system' if this is applicable to the products concerned. The different
options are listed in Annex I to this brochure.
Post-Market
Surveillance
The
manufacturer's responsibility for product quality continues after
utilization of the product has started. The directive therefore
stipulates that the manufacturer shall implement and maintain a
systematic procedure based on which the experiences gained with products
during their utilization can be evaluated, so as to be able to make
improvements to these products where necessary.
Vigilance
System
The
manufacturer shall notify the competent authorities of any incidents
involving products in use that have or could have seriously compromised
the patient's or user's health.
Furthermore,
the manufacturer shall notify the competent authorities when, following
complaints or incidents, products of similar type are systematically
taken from the market. The quality system must therefore incorporate
procedures relating to these activities.
THE NOTIFIED
BODY
A
so-called Notified Body plays an important role in the procedure for
obtaining a CE marking for devices in classes IIa, IIb and III and for
class I products placed on the market in sterile condition and/or with a
measuring function.
The
Notified Body is a certification organization designated by the national
authority of any EC country, which is authorized to conduct conformity
assessments in accordance with the procedures in the directive. The
Notified Body offers its services to the manufacturer with the aim to
have the manufacturer's product put on the European market in conformity
with the directives. The Notified Body performs its activities in a
competent, neutral, independent and non-discriminatory manner.
Conformity
Assessment Procedures
Before
being permitted to apply a CE marking, the manufacturer must demonstrate
that his product satisfies the essential requirements of the directive
and that he has an operational vigilance system for it. Moreover, one of
the procedures specified in the directive is to be observed. The
selection of a procedure depends among others on the classification of
the product.
For
products to be placed on the market in sterile condition, the notified
body will conduct an audit of the sterilization process, regardless of
the product classification, in order to verify whether the implemented
quality system ensures conformity of the sterilization process to the
relevant harmonized standards.
Class I
products:
In principle
Module A applies unless the product needs a sterile condition or has a
measuring function (in which case Modules Aa apply).
Class IIa and
IIb
products:
In principle
Module H applies, or, as an alternative, the Modules B and D, B and E or
B and F apply. With Class IIa products an EU-Type Examination
certificate (Module B) need not be issued.
Class III
products:
In principle
Module H applies, or, as an alternative, the Modules B and D or B and F
apply.
Attached
(p.t.o.) are block-diagrams, generally indicating the procedures.
CEBEC'S
METHOD OF OPERATION
The
philosophy of the CEBEC Medical Certification System is based on the
principle that the manufacturer's interests are best served if, within
the provisions of the directive, he has maximum freedom to develop new
products or modify existing ones.
Based
on this principle CEBEC Medical has developed a certification system
that considerably simplifies the certification procedures by means of
clustering. In the long term this approach, which is based on control of
the manufacturer's implemented quality system, saves time and money by
preventing the unnecessary issue of new CE certificates and/or
modification of existing certificates.
CEBEC
is a notified body (0649) for the 93/42/EEC directive.
Information
For
further information please contact:
CEBEC
Avenue. Fr. Van Kalken 9A, b. 1
B-1070 Brussels (Belgium)
Phone: +32 2 556 00 20
Fax: +32 2 556 00 36
CEBEC provides market access
|